STANDARD OPERATING PROCEDURE | ||
Title | : | Procedure for preparation of disinfectant & cleaning of microbiology laboratory |
- OBJECTIVE
- To lay down a procedure for preparation of disinfectant & cleaning of microbiology laboratory.
- SCOPE
- This SOP is applicable for disinfectant & cleaning of microbiology laboratory.
- RESPONSIBILITY
- Officer / Executive-QCM – responsible for Implementation of SOP.
- ACCOUNTABILITY
- Manager – Quality Control
- PROCEDURE
Safety Precaution:
- Use proper safety apparel such as gowns, rubber hand gloves and safety goggles during preparation and use of disinfectant.
- While preparing disinfectant solutions always add disinfectant to the water and not water to disinfectant.
- Prepare quantities of disinfectant solutions sufficient for one-day use.
- Disinfectant solutions:
- Dettol 2.5%
- Savlon 2.5 %
- Virosil 20%
- Procedure for preparation of disinfectant solution:
- Dettol (For 1 lit) 2.5%
- Take 500 ml of purified water in suitable sterile container. Add 25 ml of Dettol and mix. Make-up the volume to 1000 ml with purified water.
- Savlon (for 1 lit) 2.5%
- Take 500 ml of purified water in suitable sterile container. Add 25 ml of Savlon and mix. Make-up the volume to 1000 ml with purified water.
- IPA (for 1 Liter) 70%
- Take 300 ml of purified water in suitable sterile container. Add 700 ml of IPA & make-up to 1000 ml. Mix well.
- Virosil 20%
- Prepare the virosil solution as per manufacturer direction. Use the solution within 14 days.
- General instructions:
- Before start of day work perform the cleaning operation in classified and non classified area daily.
- Classified Area:
- Microbiology Testing Area and Airlocks
- Sterility Testing Area and Airlocks.
- Incubation area.
- Media Preparation Area.
- Non classified area:
- Wash and destruction area.
- Cleaning of Classified Area:
- Cleaning of classified area shall be done in following order.
- Sterility room → Airlock –4 → Airlock – 3→Airlock – 2 and then Airlock-1
- All disinfectant solutions shall be prepared freshly.
- Use sterile non-shedding dusters for cleaning of classified area.
- Rotate the disinfectant solution weekly.
- Record the preparation of disinfectant .
- Daily cleaning procedure:
- Clean the classified area in the following order
- Ceiling and fixtures→ Walls→ Top and side surface of LAF Bench →Floors.
- Cleaning of Ceiling, fixtures, walls, top and side surface of LAF Bench shall be done with non shedding dusters.
- Cleaning of floor shall be done using a nylon brush / non-shedding duster.
- Subsequently mop the floor using non-shedding wiping mops dipped in the disinfectant.
- Cleaning of Non Classified Area:
- All disinfectant solutions shall be prepared freshly
- Rotate the disinfectant solution weekly
- Record the preparation of disinfectant .
- Daily cleaning procedure:
- Clean the floor using a nylon brush / non-shedding duster
- Subsequently mop the floor using non-shedding wiping mops dipped in the disinfectant
- Maintain records in Cleaning record of Non Classified Area .
- Frequency:
- Ceiling and fixture – Weekly
- Walls – Weekly
- Floor – Daily
- LAF – Daily
- ABBREVIATION
- SOP – Standard operating procedure
- IPA – Iso Propyl Alcohol
- WFI – Water for Injection
- FORMAT
Sr. No. | Title | Format No. |
Micro Lab cleaning record | ||
Format for disinfectant preparation & filtration record | ||
Isopropyl Alcohol Status Label | ||
Sanitization Schedule for week in month |
DISTRIBUTION LIST
Sr. No. | Copy No. | Department |
Quality Assurance | ||
Quality Control-Microbiology |