Non-viable particle count in classified

Non-viable particle count in classified

To establish the written procedure for monitoring of non-viable particle count in the classified Area S mall Volume Parenteral section.

DEFINITIONS

Clean Area:  An area in which defined environmental control of particulate & microbial contamination, is constructed & used in such a way as to reduce the introduction, generation,& retention of contaminants within the area.

Particle Counter: A particle counter is an instrument that detects & counts physical particles. The nature of particle counting is based upon either light scattering, light obscuration, or direct imaging. A high-intensity light source is used to illuminate the particles as they pass through the detection chamber. Recorded video analyzed by computer software to measure the particle attributes.

A classified/Controlled Area is an enclosed environment or room with flawless control over particulate contamination. More especially the area has a controlled contamination level, which is specified regarding the number of particles for every cubic meter, for a specific particle size.

ABBREVIATION

SVP: Small Volume Parenteral

IPA: Isopropyl Alcohol

USP : United State Pharmacopeia

FDA: Food & Drug Administration

HEPA: High-Efficiency Particulate Air

NVPC: Non-viable particle counter

ISO: International Organization of Standardization.

RESPONSIBILITY

Originating department shall be responsible:

Preparation of the SOPs.

To follow the procedure & content of SOP.

Execution & updation of required documents as per SOP.

The head of Originating department/designee shall be responsible for:

  • To provide the training of SOP to the concerned person.
  • Implementation & compliance of procedure mentioned in SOP.

The Quality Assurance department shall be responsible for:

  • To verify that all content of SOP is relevant to the concerned equipment & process.
  • To verify that all the instructions are followed properly as per the given SOP.
  • To verify that relevant records/log books are maintained as per the defined system.

Head – Quality Assurance/ Designee shall be accountable for:

  • To approve new or revised SOP.
  • To ensure the implementation of the defined system.

The plant Head shall be accountable for:

  • To approve new or revised SOP.
  • To ensure the implementation of the defined system.

PROCEDURE:

SAFETY/PRECAUTION/EHS

  • Never change the location while the particle counter is in operation.
  • Keep the particle counter in a dry place & clean place.
  • The chemist/operator should follow all instructions of clean room behavior during the operation of the particle counter in mfg. room & filling room.
  • Cleaning shall be avoided during the monitoring of particle count.

MONITORING OF NON-VIABLE PARTICLE COUNT

Non-viable particles are to be monitored with NVPC (Non-viable particle counter) instrument as per SOP.

The purpose of non-viable particle monitoring in clean areas is to verify the environment meets the specification prior to operation or exposure of sterile products in the environment.

The limit of non-viable particle count in the area depends upon the classification of the area.

A number of sampling points, location selection, clean zone classification determination & quality of data will be generated as per ISO 14644-1 guidelines.

The minimum no. of sampling locations will be decided as per the following table of ISO 14644-1:2015 guidelines.

Area of cleanroom(m2) less than or equal to Minimum number of sample locations to be tested (NL)
2 1
4 2
6 3
8 4
10 5
24 6
28 7
32 8
36 9
52 10
56 11
64 12
68 13
72 14
76 15
104 16
108 17
116 18
148 19
156 20
192 21
232 22
276 23
352 24
436 25
500 26
1000 27
>1000 Equation A

 The area detail of the manufacturing & filling room for NVPC is as follows.                     

Sr. no. Name of Location ISO Class Area of Location (m2) Minimum no. of Sampling location
1 Manufacturing Room (I) 7 14.883 06
2 Manufacturing Room (II) 7 18.225 06
3 Filling Room 7 25 07

   

Sampling location in the area preferred at corners, near the doors & near equipment or based on air flow pattern or as per our requirement.

Clean rooms & Zones are classified as per standard ISO 14644-1, WHO TRS-961 & Schedule M Guidelines as per the following table 1, 2 & 3.

TABLE 1 (as per ISO 14644-1)

CLASS MAX. PARTICLE Per m³ as per ISO 14644-1
0.5 microns        5.0 microns
ISO-5

ISO-6

ISO-7

ISO-8

ISO-9

3520

35200

352000

3520000

35200000

*

293

2930

29300

293000

* To express an airborne concentration of 29 particles/m3 in the particle size range ≥ 5 µm in ISO Class 5 based on the use of LSAPC, the designation would be “ISO M (29; ≥ 5 µm); LSAPC”

 

 TABLE 2 (As per WHO TRS-961) 

Maximum permitted number of particles per mgreater than or equal to the tabulated size
At Rest In Operation
GRADE 0.5 µm 5.0 µm 0.5 µm 5.0 µm
A

B

C

 

D

3520

3520

352000

3520000

20

29

2900

29000

3520

352000

3520000

Not Defined

20

2900

29000

Not Defined

 

TABLE 3 (As per Schedule M) 

 

GRADE

Maximum permitted number of particles per mgreater than or equal to the tabulated size
At Rest In Operation
0.5 µm 5.0 µm 0.5 µm 5.0 µm
A

B

C

 

D

3520

3520

352000

3520000

20

29

2900

29000

3520

352000

3520000

Not Defined

29

2900

29000

Not Defined

 

NVPC in the manufacturing & filling area is to be measured with a non-viable particle counter (Make: Met-One) & flow rate of it is 100 LPM.

Transfer the instrument to the corresponding room through a dynamic pass box. Operate the instrument as per SOP).

For ISO class 7(mfg. room & filling room) the sampling time will be 1 minute to sample 2 liter of air & for ISO class 5 (FFS machine filling zone/SFS or under LAF zone) the sampling time will be 10 minutes to sample 1000 liters of air. Attach the print report of NVPC in BMR.

Frequency

For (SFS or LFR) ISO class 5 of filling machine: during startup of filling operation.  

For (manufacturing room) ISO class 7: before & after every batch processing.

For (Filling room) ISO class 7: before starting and completion of every batch filling operation.

After the major breakdown in the filling room or FFS machine, as the machine starts again, NVPC will be monitored for the filling zone & filling room.

REFERENCE

WHO-TRS 961 guideline

FDA and schedule M

ISO-14644