Procedure for autoclave validation
STANDARD OPERATING PROCEDURE | |||
Title | : | Procedure for autoclave validation | |
Department | : | Quality Control-Microbiology | |
- OBJECTIVE
- To lay down a procedure for autoclave validation.
- SCOPE
- This SOP is applicable for autoclave validation in microbiology laboratory .
- RESPONSIBILITY
- Officer / Executive-QC – responsible for Implementation of SOP.
- ACCOUNTABILITY
- Manager – Quality Control
- PROCEDURE
- Method-A:
- Clean and load the autoclave for sterilization with accessories and glassware’s.
- Prepare the required number of (10ml) tubes of Soya bean casein digest medium.
- Take require nos. of strips of spores of Geobacillus stearothermophilus ATCC 7953.
- Put in petriplates and label it, arrange the petiplates at different location no.
- Keep one strip unexposed for positive control.
- Run the cycle at 15 lbs. pressure and 121o C temperature for 15 min.
- After completion of cycle remove all tubes and Petri dishes.
- Mark location no.,date on all the SCDM tube .
- Transfer strips aseptically to respective labelled tubes.
- Transfer one unexposed strip to one SCDM tube and label it as positive control.
- Keep one SCDM tube uninoculated and label it as negative control.
- Incubate all the tubes at 55o C ±2o C for 7days
- At the end of 7 days all the the inoculated SCDM tubes should show no growth.
- Negative control tube should show no growth.
- Positive control tube should show growth with in 72 hrs.
- The above observations will demonstrate satisfactory validation of the autoclave.
- Record the observation in Format-I.
- Method-B
- Validation is done by the bioindicator system consists of ampoule containing spores of Geobacillus stearothermophillus ATCC 7953.
- These ampoules are stored under refrigerator when not in use. On the day of validation
- Take three ampoules and allow them to come at room temperature.
- Load the autoclave and place the ampoule in different locations in autoclave within glass beakers to avoid accidental breakage of ampoule during sterilization.
- Run the sterilization cycle as per procedure. (validated sterilization condition 121°C for 15 min at 15 PSI.)
- Unload the chamber and remove the ampoule, mark them and incubate at 55°C ±2°C for 48 hour.
- The third non sterile unexposed ampoule is incubated along with the sterilized ampoule as a Positive control.
- At the end of 48 hour days.
- Positive control – Original purple colour gradually changes to deep yellow with turbidity.
- Exposed ampoule – The Original purple colour remains as it is at the end of incubation period.
- The above observations will demonstrate satisfactory validation of the autoclave.
- Record the result in Format-I
- ABBREVIATION
- SOP – Standard operating procedure
- IPA – Iso Propyl Alcohol
- SCDM – Soybean Casein Digest Medium
- FORMAT
Sr. No. | Title | Format No. |
Log sheet for Autoclave validation Record | ||
Log Sheet for Sterilization load no. |
- DISTRIBUTION LIST
Sr. No. | Copy No. | Department |
Quality Assurance | ||
Quality Control-Microbiology | ||
V.P.- Plant & Operations |
Format -I
- Equipment Detail:
Name of Equipment | Vertical Autoclave |
Equipment ID No |
- Detail of Test:
Sr.No | Description | Detail |
Temperature/pressure to be achieved | 121°C/15lbs | |
Biological indicator used | Strip/Ampoule | |
Name of the test organism | Bacillus stearothermophilus ATCC-7953 | |
Total population of strip/Ampoule | ||
D value. | ||
Media to be use | Soybean Casein Digest broth | |
Incubation temperature | 55°C to 60°C | |
Duration of Incubation |
- Details of Validation:
Details of last Validation: | |
Validation On | |
Validation Due on | |
Report No. | |
Detail of current Validation: | |
Date of Validation | |
Validation Due on | |
Report No. |
- Details of Observation:
Sr.No | Parameters | Observation | Observed by |
Temperature/Pressure achieved | |||
Name of the test organism | |||
Media used | |||
Incubation temperature | |||
Duration of Incubation | |||
Lot number of Strip/Ampoule | |||
Manufacturer of strip/Ampoule |
- Acceptance Criteria:
There should not be any growth in form of turbidity in any of the tubes except Positive control.
- Corrective action if observation does not meets the requirements:
Corrective action taken: |
Observation after corrective action:
|
- Conclusion:
|
- Approval:
OBSERVATION SHEET
Sr. No | Location | Day 1 | Day 2 | Day 3 | Day 4 | Day 5 | Day 6 | Day 7 | Remark |
Negative Control | |||||||||
Positive Control | |||||||||
Observed by |
- Format -II
Log Book of Sterilization(Autoclave) Load Number
Instrument I.D.: ______________________
Sr.No. | Date | Name of Sample | Load Number | Done by | Checked by | Remark |
| ||||||
| ||||||
|